Biotronik

14,000,000 Leading Edge Experts on the ideXlab platform

Scan Science and Technology

Contact Leading Edge Experts & Companies

Scan Science and Technology

Contact Leading Edge Experts & Companies

The Experts below are selected from a list of 4341 Experts worldwide ranked by ideXlab platform

Marc Bentele - One of the best experts on this subject based on the ideXlab platform.

T Meinertz - One of the best experts on this subject based on the ideXlab platform.

  • Effects of a thin-sized lead body of a transvenous single coil defibrillation lead on ICD implantation. Kainox RV Study Group.
    Pacing and clinical electrophysiology : PACE, 2020
    Co-Authors: A Schuchert, T Meinertz
    Abstract:

    In the interest of patients receiving implantable cardioverter defibrillators (ICDs), the clinical benefits of newer and thinner transvenous defibrillation leads have to be determined. The aims of this study were to evaluate the ICD procedure duration and the frequency of lead dislocation at the 3-month follow-up of a new defibrillation lead with a thin-sized lead body and its conventional-sized predecessor. The thin-sized single coil defibrillation lead (Kainox RV, Biotronik; lead body 6.7 Fr) was implanted in 61 patients and the conventional-sized defibrillation lead (SPS, Biotronik; lead body 7.8 Fr) in 60 patients. Both leads were connected to a left-sided, prepectorally implanted Phylax ICD (Biotronik) with active housing. The lead implantation time and total procedure duration were determined. Lead implantation time was defined as the time from lead insertion to the end of the pacing measurements. The total procedure duration spanned skin incision to closure. The incidence of lead repositioning during the lead implantation time and during ventricular fibrillation conversion testing was also assessed. The frequency of lead dislocations was recorded at the 3-month follow-up. Mean lead implantation time and total procedure duration of the thin-sized lead (23 +/- 22 minutes 76 +/- 37 minutes) were not statistically different from the time needed for the conventional-sized lead (22 +/- 20 minutes 81 +/- 34 minutes). The number of lead repositionings during the lead implantation time was similar (thin-sized lead: 1.4 +/- 2.4; conventional-sized lead: 1.1 +/- 1.9). An additional lead repositioning was not necessary during ventricular fibrillation conversion testing in 93.4% of the patients with thin-sized and in 94.4% with conventional-sized leads (not significant). At the 3-month follow-up, there were four (6.6%) lead dislocations in the thin-sized and four (6.7%) in the conventional-sized lead group. In conclusion, the down-sized lead body of the new defibrillation lead influenced neither ICD procedure duration nor the incidence of lead dislocation during follow-up.

  • Effects of a Thin-Sized Lead Body of a Transvenous Single Coil Defibrillation Lead on ICD Implantation
    Pacing and Clinical Electrophysiology, 2000
    Co-Authors: A Schuchert, T Meinertz
    Abstract:

    SCHUCHERT, A., et al.: Effects of a Thin-Sized Lead Body of a Transvenous Single Coil Defibrillation Lead on ICD Implantation. In the interest of patients receiving implantable cardioverter defibrillators (ICDs), the clinical benefits of newer and thinner transvenous defibrillation leads have to be determined. The aims of this study were to evaluate the ICD procedure duration and the frequency of lead dislocation at the 3-month follow-up of a new defibrillation lead with a thin-sized lead body and its conventionalsized predecessor. The thin-sized single coil defibrillation lead (Kainox RV, Biotronik; lead body 6.7 Fr) was implanted in 61 patients and the conventional-sized defibrillation lead (SPS, Biotronik; lead body 7.8 Fr) in 60 patients. Both leads were connected to a left-sided, prepectorally implanted Phylax ICD (Biotronik) with active housing. The lead implantation time and total procedure duration were determined. Lead implantation time was defined as the time from lead insertion to the end of the pacing measurements. The total procedure duration spanned skin incision to closure. The incidence of lead repositioning during the lead implantation time and during ventricular fibrillation conversion testing was also assessed. The frequency of lead dislocations was recorded at the 3-month follow-up. Mean lead implantation time and total procedure duration of the thin-sized lead (23 ± 22 minutes 76 ± 37 minutes) were not statistically different from the time needed for the conventional-sized lead (22 ± 20 minutes 81 ± 34 minutes). The number of lead repositionings during the lead implantation time was similar (thin-sized lead: 1.4 ± 2.4; conventional-sized lead: 1.1 ± 1.9). An additional lead repositioning was not necessary during ventricular fibrillation conversion testing in 93.4% of the patients with thin-sized and in 94.4% with conventional-sized leads (not significant). At the 3-month follow-up, there were four (6.6%) lead dislocations in the thin-sized and four (6.7%) in the conventional-sized lead group. In conclusion, the downsized lead body of the new defibrillation lead influenced neither ICD procedure duration nor the incidence of lead dislocation during follow-up.

Andreas Kucher - One of the best experts on this subject based on the ideXlab platform.

  • Left ventricular inhibition during cardiac resynchronization caused by sensed left-sided ventricular premature complexes.
    Pacing and Clinical Electrophysiology, 2020
    Co-Authors: S. Serge Barold, Andreas Kucher
    Abstract:

    : A recently published case documented left ventricular (LV) inhibition of a Boston Scientific device by a premature complex (VPC) that was undetected by the right ventricular channel. We have observed a similar response in two patients with a Biotronik cardiac resynchronization device also designed with LV sensing. A spurious response simulating that of the two true cases was also observed in a third patient with a defective LV lead which created isolated false signals. The responses of the Biotronik devices were identical to that of the previously reported findings with the Boston Scientific device. The observations provide insight into the timing function of cardiac resynchronization devices designed with LV sensing.

  • Interruption of cardiac resynchronization therapy by atrial premature complexes
    Journal of Electrocardiology, 2017
    Co-Authors: S. Serge Barold, Andreas Kucher
    Abstract:

    Abstract Biotronik devices used for cardiac resynchronization therapy (CRT) combined with defibrillation function (CRT-D) are capable of left ventricular (LV) sensing. Under certain circumstances, LV sensing may cause loss of CRT. The third generation of the Biotronik i-family CRT-Ds enables the recording of event-triggered tracings of the electrogram particularly those involving “CRT pacing interrupt” episodes. We report three cases of a sudden “CRT pacing interrupt” initiated by an atrial premature complex. This was caused by realignment of the LV timing cycles induced by the APCs whereupon LV pacing was inhibited and a self-perpetuating desynchronization process was initiated. In all cases it is the repeated occurrence of LV sensed events that prevents the emission of LV paced events because it displaces the LV upper rate interval from its normal position. Prevention of desynchronization requires programming an LV upper rate faster than the maximum sensor-driven rate or right ventricular upper rate.

  • The Analog Blanking Period of Implantable Cardiac Rhythm Devices.
    Pacing and Clinical Electrophysiology, 2017
    Co-Authors: S. Serge Barold, Andreas Kucher
    Abstract:

    Background Analog blanking periods (BPs) that hold down the display of electrograms (EGMs) in cardiac rhythm devices have received much less attention than the well-known digital BPs which do not influence the EGM display. In Biotronik devices (Biotronik GmbH, Berlin, Germany), when a paced event initiates an analog BP in one chamber (right atrium, right ventricle [RV], or left ventricle [LV]), an identical cross-chamber analog BP starts in the other two chambers. Methods All clinical observations were recorded from patients with Biotronik devices. The effect of the atrial cross-chamber analog BP initiated by a ventricular paced event on the atrial EGM was studied in the records of seven patients with frequent paroxysmal atrial flutter to determine the effect of critically timed RV paced event (RVp) or LV paced event (LVp) upon the atrial EGM. The effect of atrial pacing triggering cross-chamber analog BPs in the RV and LV channels on the RV and LV EGMs was also investigated in cases of conducted supraventricular beats and ventricular premature complexes. The effect of a triggered LVp initiating a cross-chamber analog BP in the RV channel on the EGM of a sensed RV sensed event was evaluated in 10 cases. Simulation studies were also performed to verify the clinical observations. Results Patients with atrial flutter showed intermittent truncation or deformity and even disappearance of the atrial signals due to an atrial cross-chamber analog BP initiated by RVp and/or LVp. Three patients demonstrated deformity of the signal shape of ventricular premature beats falling within a ventricular cross-chamber analog BP initiated by right atrial paced event (RAp). We found only one case of a supraventricular QRS complex trapped in a ventricular cross-chamber analog BP initiated by RAp. All the recordings of LVp triggering upon RVs revealed a variety of RV signal deformities occasionally with preservation of the terminal part of the RV signal. Simulation studies confirmed the effect of the analog BP on the atrial and the ventricular EGMs. Conclusion The analog BP of Biotronik devices may cause truncation or deformity of atrial and ventricular signals and the occasional disappearance of an atrial signal during atrial flutter. These effects must not be interpreted as device malfunction. In the clinical evaluation of the EGM curves, the effects of the analog BPs have to be carefully considered in order to avoid possible misinterpretation.

  • Understanding the Timing Cycles of a Cardiac Resynchronization Device Designed with Left Ventricular Sensing
    Pacing and Clinical Electrophysiology, 2014
    Co-Authors: S. Serge Barold, Andreas Kucher
    Abstract:

    : Some devices used for cardiac resynchronization therapy (CRT) can sense from the left ventricular (LV) lead as in Biotronik CRT devices (Biotronik GmbH, Berlin, Germany), whose special LV timing cycles form the basis of this report. LV sensing (LVs) was designed to prevent competitive pacing outside the LV myocardial absolute refractory period. LVs works by inhibiting the release of an LV pacemaker stimulus (LVp) in the vulnerable period of the LV during a programmable period. LVs with stored LV electrograms may also provide recordings of diagnostic value in tachyarrhythmias. LVs has added a new dimension to the evaluation of the function of CRT devices, because it is manifested by unfamiliar timing cycles. In this respect, Biotronik devices can initiate an LV upper rate interval (URI) upon sensing a right-sided event when LVs is turned off. An inhibited LVp can also initiate an LVURI. The LVURI should generally be programmed to a relatively short duration and shorter than the right ventricular URI to prevent a special form of desynchronization arrhythmia sustained by LVs. This arrhythmia is characterized by recurring delayed LVs events in sequences associated with RV pacing followed by LVs events with loss of LVp.

  • Rise in ICD shock impedance: lead fracture or death?
    Pacing and Clinical Electrophysiology, 2012
    Co-Authors: Roland X. Stroobandt, Andreas Kucher, Frederic E. Van Heuverswyn, S. Serge Barold
    Abstract:

    BACKGROUND: Remote monitoring allows for interrogation and extensive data retrieval of implantable cardioverter-defibrillators (ICDs). Data on ICD parameters at the time of death and afterwards are limited. The purpose of this retrospective study was to examine the changes in lead impedances of ICDs at the time of death and afterwards. METHODS: A total of 37 Biotronik (SE & CO. KG, Berlin, Germany) ICDs (20 ICD-cardiac resynchronization therapy, 16 dual-chamber ICDs, and one single-chamber ICD), retrieved after death, were interrogated. Stored intracardiac electrograms were analyzed to determine the cause of death. Impedance trend curves of shock and pacing lead impedances were analyzed and correlated retrospectively with the reported time of death. The influence of cold exposure on lead impedances was tested in three other single-chamber Biotronik ICDs. RESULTS: Of 37 patients, the cause of death was due to ventricular tachyarrhythmias in 21 patients. In 12 patients, death was not arrhythmia-related. In four patients, the cause of death could not be determined due to overwriting of the episodes at the time of death. A significant increase of shock and pacing lead impedances was observed in the postmortem days (P < 0.001 for all lead impedances). All lead impedance values increased significantly within the first postmortem day (P < 0.001 for all lead impedances). Cold exposure decreased shock lead impedance but did not affect pacing lead impedance. CONCLUSION: Postmortem analysis of ICDs allows tracking of lead impedance changes, which correlate with the day of death. The rise in postmortem impedances should not be interpreted as contributing to the mode of death.

Soeren Mattke - One of the best experts on this subject based on the ideXlab platform.

A Schuchert - One of the best experts on this subject based on the ideXlab platform.

  • Effects of a thin-sized lead body of a transvenous single coil defibrillation lead on ICD implantation. Kainox RV Study Group.
    Pacing and clinical electrophysiology : PACE, 2020
    Co-Authors: A Schuchert, T Meinertz
    Abstract:

    In the interest of patients receiving implantable cardioverter defibrillators (ICDs), the clinical benefits of newer and thinner transvenous defibrillation leads have to be determined. The aims of this study were to evaluate the ICD procedure duration and the frequency of lead dislocation at the 3-month follow-up of a new defibrillation lead with a thin-sized lead body and its conventional-sized predecessor. The thin-sized single coil defibrillation lead (Kainox RV, Biotronik; lead body 6.7 Fr) was implanted in 61 patients and the conventional-sized defibrillation lead (SPS, Biotronik; lead body 7.8 Fr) in 60 patients. Both leads were connected to a left-sided, prepectorally implanted Phylax ICD (Biotronik) with active housing. The lead implantation time and total procedure duration were determined. Lead implantation time was defined as the time from lead insertion to the end of the pacing measurements. The total procedure duration spanned skin incision to closure. The incidence of lead repositioning during the lead implantation time and during ventricular fibrillation conversion testing was also assessed. The frequency of lead dislocations was recorded at the 3-month follow-up. Mean lead implantation time and total procedure duration of the thin-sized lead (23 +/- 22 minutes 76 +/- 37 minutes) were not statistically different from the time needed for the conventional-sized lead (22 +/- 20 minutes 81 +/- 34 minutes). The number of lead repositionings during the lead implantation time was similar (thin-sized lead: 1.4 +/- 2.4; conventional-sized lead: 1.1 +/- 1.9). An additional lead repositioning was not necessary during ventricular fibrillation conversion testing in 93.4% of the patients with thin-sized and in 94.4% with conventional-sized leads (not significant). At the 3-month follow-up, there were four (6.6%) lead dislocations in the thin-sized and four (6.7%) in the conventional-sized lead group. In conclusion, the down-sized lead body of the new defibrillation lead influenced neither ICD procedure duration nor the incidence of lead dislocation during follow-up.

  • Effects of a Thin-Sized Lead Body of a Transvenous Single Coil Defibrillation Lead on ICD Implantation
    Pacing and Clinical Electrophysiology, 2000
    Co-Authors: A Schuchert, T Meinertz
    Abstract:

    SCHUCHERT, A., et al.: Effects of a Thin-Sized Lead Body of a Transvenous Single Coil Defibrillation Lead on ICD Implantation. In the interest of patients receiving implantable cardioverter defibrillators (ICDs), the clinical benefits of newer and thinner transvenous defibrillation leads have to be determined. The aims of this study were to evaluate the ICD procedure duration and the frequency of lead dislocation at the 3-month follow-up of a new defibrillation lead with a thin-sized lead body and its conventionalsized predecessor. The thin-sized single coil defibrillation lead (Kainox RV, Biotronik; lead body 6.7 Fr) was implanted in 61 patients and the conventional-sized defibrillation lead (SPS, Biotronik; lead body 7.8 Fr) in 60 patients. Both leads were connected to a left-sided, prepectorally implanted Phylax ICD (Biotronik) with active housing. The lead implantation time and total procedure duration were determined. Lead implantation time was defined as the time from lead insertion to the end of the pacing measurements. The total procedure duration spanned skin incision to closure. The incidence of lead repositioning during the lead implantation time and during ventricular fibrillation conversion testing was also assessed. The frequency of lead dislocations was recorded at the 3-month follow-up. Mean lead implantation time and total procedure duration of the thin-sized lead (23 ± 22 minutes 76 ± 37 minutes) were not statistically different from the time needed for the conventional-sized lead (22 ± 20 minutes 81 ± 34 minutes). The number of lead repositionings during the lead implantation time was similar (thin-sized lead: 1.4 ± 2.4; conventional-sized lead: 1.1 ± 1.9). An additional lead repositioning was not necessary during ventricular fibrillation conversion testing in 93.4% of the patients with thin-sized and in 94.4% with conventional-sized leads (not significant). At the 3-month follow-up, there were four (6.6%) lead dislocations in the thin-sized and four (6.7%) in the conventional-sized lead group. In conclusion, the downsized lead body of the new defibrillation lead influenced neither ICD procedure duration nor the incidence of lead dislocation during follow-up.